Abstract. A force platform is a technical method
of quantitatively assessing balance indirectly. The use of force
platforms in physiotherapy departments has become more prominent
over the last few years. However, the main drawback in the use
of force platforms is the lack of comprehensive calibration procedures,
which casts doubt on the results obtained with these systems.
Existing calibration tests are limited to testing the spatial
accuracy of the force platform. This paper describes a comprehensive
quality control test procedure, which was developed. It is proposed
that the developed quality control test procedure could be used
to test all types of force platform and it includes a description
of how the tests should be carried out, the frequency with which
they should be carried out and the expected performance for each
of the tests as recommended for the most part by the Association
Française de Posturologie.
Force
platforms are technical methods of assessing balance and as such
should have calibration results to verify the accuracy and precision
of their clinical results. A significant deficiency in the use
of force platforms is the lack of a comprehensive quality control
test procedure which could be used to assess the accuracy and
precision of the force platform measurements on a month-to-month
or year-to-year basis (Andres and Anderson 1980). The lack of
such a quality control procedure casts doubt on the reliability
of the various measured indices of balance described below, and
also makes it impossible to compare results from month-to-month
within the same clinic and between different clinics, even for
the same type of force platform. It is very important to perform
quality control tests so that it can be established whether the
performance of the force platform is within the specifications.
A force platform
is an indirect technical method of measuring balance or sway,
where the force exerted by the body on the ground is measured.
From this measurement the person's centre of pressure (COP) is
determined, and this is then related to their centre of gravity
(COG) and ultimately to their balance. This is due to the relationship
between these three parameters: the movement of a person's centre
of gravity is a measure of their balance, and during quiet standing
a persons centre of pressure can be approximated to their centre
of gravity (Winter 1995). Force platforms measure the COP using
force or pressure transducers, which produce an electrical signal
proportional to the applied force.
|
Figure 1. Standard force platform. |
There are many different types of force or pressure transducer in use: strain gauges and piezoelectric, piezoresistive and capacitive transducers (Winter 1990). A standard force platform is composed of a flat top plate supported by three or four force or pressure transducers. The location of a person's COP is determined by the relative forces experienced at each of these corner transducers. Figure 1 is a diagram of a standard force platform.
An analysis of the motion of the COP provides detailed quantitative measures of a person's postural stability, called indices of balance. Examples of some of these indices of balance are the following: |
Balance disorders
can result from a loss of vestibular information, Menier's disease
or vertigo; from damage to the CNS or the cerebellum, stroke or
Parkinson's disease and also from damage to the musculoskeletal
system. Different indices of balance are sensitive to the presence
of different types of balance disorder (Guerts et al. 1993,
Kilbum et al. 1994, Hufschmidt et al. 1980). The
changes in the indices of balance due to impaired balance are
not very large, 5 mm for a healthy volunteer and 7.1 mm for a
patient with moderate Parkinson's disease, a difference of only
2.1 mm (Browne 1999). Therefore, the force platform measurements
of the indices of balance need to be both accurate and precise.
Numerous
investigators have carried out calibration tests on force platforms;
however, these calibration tests have been mostly limited to assessing
only the spatial accuracy of the force platform and the sources
of error causing the inaccuracy (Bobbert and Schamhardt 1990,
Mita et al. 1993, Granat et al. 1990, Hall et
al. l996, Starck et al. 1993). Each of the different
investigators identified different sources of error for this spatial
inaccuracy: excessive nonlinearity of the transducer response;
different offset voltages of each of the transducers and electronic
noise. Hall et al. (1996) suggested that a large source
of error in the spatial accuracy of the force platform was the
installation of the force platform and recommended that a calibration
be performed at installation to minimise these errors. This calibration
was limited to only determining the output and the crosstalk of
each of the transducers. However, despite a knowledge of the errors
which cause spatial inaccuracy no calibration procedure bas been
developed to include the testing of the different parameters which
contribute to the errors in spatial accuracy, such as the level
of nonlinearity of the transducers or electronic noise in the
analogue-to-digital converter signal of the force platform. Therefore,
a quality control test procedure for characterising force platforms
was developed. This quality control test procedure includes tests
for all of the important aspects of the performance of force platforms,
and the recommended performance criteria are for the most part
those recommended by the Association Française de Posturologie
to standardise the construction of force platforms (Bizzo et
al. 1985). The Association Française de Posturologie
recognised the need for standard construction specifications for
force platforms to be developed and, therefore, outlined specifications
for construction of a force platform with a spatial accuracy of
1 mm and a natural frequency greater than 20 Hz. Despite its obvious
importance, no recommendation for precision was made by the Association
Française de Posturologie and, also, it has not been investigated
by other investigators.
This quality
control test procedure investigates the performance of force platforms
from a static and a dynamic perspective. The static performance
of the force platform is governed by its spatial accuracy and
precision while the dynamic performance is governed by its frequency
response. The spatial accuracy of the force platform is affected
by the electronic noise and hysteresis of the system, the nonlinearity
and different offset voltages of the transducers as mentioned
earlier, while the precision of force platform is affected by
the repeatability and the temporal stability of the system. Therefore,
in order for the static performance of the force platform to be
fully investigated linearity, hysteresis, noise, repeatability
and temporal stability will be investigated as well as spatial
accuracy and uniformity. In order for the dynamic performance
of the force platform to be investigated the natural frequency
and the frequency response, modulated transfer function (MTF)
will be investigated. In this paper a quality control (QC) procedure
for all types of force platform is proposed.
2.1. Quality control test procedure
This
QC procedure includes QC tests which may be carried out on force
platforms which either have access only to the COP measurements
or have access to both the COP measurements and to the voltage
output from the analogue to digital converter for each of the
transducers. The part of the QC procedure which uses the COP measurements
will only provide general information about the force platform's
accuracy and precision and will only allow problems with nonlinearity,
hysteresis, noise, repeatability and stability of the force platform
to be identified, whereas the part of the QC procedure which uses
the output voltage from the analogue-to-digital converter can
pinpoint which transducer has excessive nonlinearity, hysteresis,
non-repeatability and instability.
All measurements should be made at room temperature (22°C)
and the temperature recorded at 30-min intervals while the QC
procedure is being carried out.
2.1.1. Static tests of the QC procedure.
Spatial
accuracy is one of the most important
parameters of a force platform's characteristics as it is essential
for a patient's treatment management that small changes over time
can be accurately detected (Berg 1989). Spatial accuracy is affected
by nonlinearity of the transducer's response to applied load;
hysteresis in the transducers; different offset voltages of the
individual transducers and electronic noise in the individual
components of the force platform. For example if the transducers
have a nonlinear response to applied load, then the COP may be
underestimated or overestimated. The point of force application
is measured using the force platform's own processing and analysis
program; therefore the same test is used regardless of whether
or not the output voltage from the analogue-to-digital converter
is accessible. This is one of the more difficult tests to perform
in the QC procedure because of the difficulty with positioning
the calibrated loads on the top plate of the system. If an accurate
co-ordinate grid is placed on the top plate of the force platform,
it will facilitate more accurate positioning of the loads on the
top plate.
A number
of different measurement techniques have been developed.
The spatial
accuracy was measured by Barrett et al. (1987) and Mita
et al. (1993) by placing different magnitudes of calibrated
loads on the co-ordinate of interest. The loads were then removed
and reapplied on the next co-ordinate of interest. This was repeated
for up to 40 co-ordinates, which gave an adequate representation
of the co-ordinates of the top plate for the range of loads between
7 and 30 kg. However, this method is time consuming and subject
to large sources of inaccuracies in stacking the loads.
The spatial
accuracy was measured by Starck et al. (1993), Bobbert
and Schamhardt (1990), Hall et al. (1996) and Gill and
O'Connor (1997) using dedicated rigs for applying loads of different
magnitudes to specific locations, which gave an adequate representation
of the co-ordinates of the top plate. A description of the construction
of one such rig is described by Gill and O'Connor (1997).
In this investigation
a translational movement system was constructed, which translated
a range of calibrated loads between 10 and 80 kg across the top
plate of the force platform, without having to keep removing and
reapplying the loads (figure 2). The accuracy of the force platform
is determined in a relative manner by the translational movement
system, the loads are moved by 2 cm and the accuracy with which
the force platform can measure this 2-cm change in position is
determined. The translational movement system consisted of a moveable
plate on which loads were stacked and then easily and accurately
moved by 2 cm (an accuracy of ±l mm). The loads were applied
to 40 different locations, which gave an adequate representation
of the co-ordinate of the top plate. The co-ordinate points measured
were those contained within the area ±10 cm from the origin
in both the x- and y-direction (Bizzo et al. 1985).
The accuracy
of the COP measurements should be accurate to within 1 mm when
a 10 kg load is placed 10 cm out from the mechanical centre of
the force platform (Bizzo et al. 1985, Barrett et al.
1987, Mita et al. 1993, Starck et al. 1993, Gill
and O'Connor 1997).
This test
should be carried out as part of acceptance testing and at a minimum
of six monthly intervals thereafter unless the stability of the
system is such that it needs to be carried out more frequently.
Uniformity is also an important parameter of the force platform's
characteristics. If the force platform's top plate has a non-uniform
response, then the measurements of a patient's COP will be affected
if the patient stands on different positions of the top plate,
or if the area of their COP increases it will not be measured
accurately.
|
Figure 2. Translational movement system. |
Uniformity is affected by nonlinearity of the transducers response to load; hysteresis; different offset voltages of the individual transducers; electronic noise in the individual components of the force platform and deformation of the top plate. The uniformity of the force platform's top plate is measured using the force platform's own processing and analysis program. The uniformity of the force platform top plate can be determined by applying a calibrated load between 10 and 40 kg at a sufficiently representative number of co-ordinate points on the top plate. |
The co-ordinate
points should be contained within the area ± l0 cm from
the origin in both the x- and y-direction (Bizzo et al.
1985). It is imperative that the load is accurately placed at
each of the co-ordinates of interest and this can be achieved
by placing a mat with the traced circumference of the base of
the load on it and then positioning the load within the trace
(figure 3).
The force
platform's top plate should have a uniform response to within
1 mm across the area contained within ± 10 cm from the
mechanical centre in both the x- and y-directions (Bizzo et
al. 1985).
|
This test should be carried out as part of acceptance testing and at a minimum of monthly intervals thereafter unless the stability of the system is such that it needs to be carried out more frequently. It is also necessary to carry out a smaller daily check on the uniformity of the force platform, in which a calibrated load between 10 and 40 kg is applied on at least five random co-ordinates, which adequately represent the top plate of the force platform. The co-ordinate points should be contained within the area ±10 cm from the mechanical centre in both the x- and y-direction (Bizzo et al. 1985). In normal operation the range of applied loads for subjects is between 40 and 160 kg and each of the transducers will have this range of loads applied to them divided by the number of transducers of the platform. |
Figure 3. Top plate of force platform with a positioning mat. |
Therefore, the
linearity of the force platform's transducers should
be determined for this range of loads.
The linearity of the force platform can be determined by applying
calibrated loads of between 5 and 100 kg to different positions
on the top plate of the force platform and recording the COP measurements
calculated. The COP measurements should not vary by more than
1 mm. Therefore, the response of the COP measurements should be
independent of the load applied (Bizzo et al. 1985).
Additional
information about the linearity of the individual transducers
can be obtained if the output voltage from the analogue-to-digital
converter (ADC) is accessible. The linearity of the transducers
can be determined by applying calibrated loads of between 5 and
100 kg directly over each of the transducers and the output voltage
response of each of the transducers recorded for the increasing
magnitude of load. The offset voltage of each of the transducers
can also be obtained. The offset voltages of each of the transducers
should be the same for all of them otherwise the COP measurements
will be inaccurate (Beppu et al. 1985, Mita et al.
1993). The transducers should have a nonlinearity < 0.1% full
scale (Bizzo et al. 1985).
This test
should be performed as part of acceptance testing and at a minimum
of yearly intervals thereafter unless the stability of the system
is such that it needs to be carried out more frequently.
In normal
operation the loads applied to each of the transducers will increase
and decrease within a few milliseconds as the subject sways to
and from the transducers.
In order
to simulate the clinical test situation a calibrated load between
20 and 40 kg should be applied to different positions on the top
plate of the force platform contained within the area ±10
cm from the origin in both the x- and y- direction and the COP
measurements recorded. Then a smaller calibrated load between
5 and 10 kg should be applied on top of the larger load and again
the COP measurement recorded. The smaller load should then be
removed and the COP measurement for the original calibrated load
recorded. This test should be repeated at least four times at
different positions contained within the area ±10 cm from
the origin in both the x- and y-direction for an accurate measurement
of hysteresis present in the system to be determined. The COP
measurements should not vary by more than 1 mm (Bizzo et al.
1985).
Additional
information about the hysteresis of the individual
transducers can be obtained if the output voltage from the ADC
is accessible. The level of hysteresis in each of the transducers
can be determined by applying a calibrated load between 10 and
40 kg directly over each of the transducers individually and the
output voltage response of each of the transducers recorded once
the output voltage becomes stable. Then a smaller additional calibrated
load between 5 and 10 kg should be applied on top of the larger
load and again the output voltage of each of the transducers recorded
once the output voltage becomes stable. The smaller load should
then be removed and the output voltage response of the transducers
for the original calibrated load recorded again once the output
voltage becomes stable. The above procedure should be repeated
at least four times so that the amount of hysteresis present in
the transducers can be determined. The transducers should have
hysteresis < 0.2% full scale (Bizzo et al. 1985).
This test
should be performed as part of acceptance testing and at a minimum
of yearly intervals thereafter unless the stability of the system
is such that it needs to be carried out more frequently.
The noise
present in the force platform's measurements can be determined
by applying a calibrated load between 10 and 40 kg to the force
platform and the calculated displacement of the COP of the static
calibrated load is representative of the amount of noise present
in the system. The displacement of the COP measurement should
be < 1 mm (Bizzo et al. 1985).
Additional
information about the contribution of the transducers and the
ADC can be obtained if the output voltages from the ADC is accessible.
The noise in the force platform's measurements can be determined
by applying a calibrated load between 10 and 40 kg directly over
each of the transducers and measuring the amount of variance in
the output voltage signal of the ADC, which corresponds to the
noise in the signal. The noise level should contribute to <
1 mm inaccuracy in the spatial accuracy of the force platform
COP measurements (Bizzo et al. 1985, Granat et al.
1990).
This test
should be performed as part of acceptance testing and at a minimum
of six monthly intervals thereafter unless the stability of the
system is such that it needs to be carried out more frequently.
Repeatability contributes to the precision
of the COP measurements: it is important that the COP measurements
are precise so that changes over time in the COP measurements
can be detected. The repeatability of the force platform can be
determined by applying a calibrated load of between 10 and 40
kg to the force platform at least 20 times in the same position
and recording the COP measurement obtained each time. It is important
that the load is accurately placed in the same position each time:
this can be achieved by using a placement mat as described in
the uniformity test procedure.
Additional
information about the repeatability of the individual transducers
can be obtained if the output voltage from the ADC is accessible.
The repeatability of the transducers can be determined by applying
a calibrated load between 10 and 40 kg directly over each of transducers
individually at least 20 times and measuring the output voltage
response of the transducers from the ADC. It is important that
the load is accurately placed in the same position each time:
this can be achieved by using a placement mat as described in
the uniformity test procedure.
The Association
Française de Posturologie made no recommendations for the
repeatability of COP measurements; however, upon consideration
of the small magnitude of difference, 2 mm, between patients with
Parkinson's disease and age-matched healthy subjects it is evident
that COP measurements should also be precise to within 1 mm, in
order to detect changes in subject's ability to balance (Browne
1999, Berg 1989, Hufschmidt et al. 1980).
This test
should be carried out as part of acceptance testing and at a minimum
of weekly intervals thereafter unless the stability of the system
is such that it needs to be carried out more frequently.
Drift and temporal stability also contribute to
the precision of the COP measurements. The drift in the force
platform measurements over a 5-30 minutes time interval can be
measured by placing a calibrated load between 10 and 40 kg on
the top plate and recording the COP measurements every 5 minutes.
Additional
information about the drift in the individual transducers can
be obtained if the output voltage from the ADC is accessible.
The drift in the transducers over a 5-30 minutes time interval
can be measured by placing a calibrated load between 10 and 40
kg over each of the transducers and recording the output voltage
response of the transducers from the ADC every 5 minutes.
The stability
of the force platform over time can be determined by applying
a calibrated load between 10 and 40 kg to the force platform and
recording the COP measurement obtained each time. It is important
that the load is accurately placed in the same position each time:
this can be achieved by using a placement mat as described in
the uniformity test procedure.
| Test |
|
Frequency test carried out |
| Linearity |
|
Installation & yearly b |
| Hysteresis |
|
Installation & yearly b |
| Noise |
|
Installation & six monthly b |
| Repeatability |
|
Installation & weekly b |
| System Drift |
|
Daily b |
| Temporal Stability |
|
Monthly b |
| Spatial accuracy |
|
Installation & six monthly b |
| Uniformity |
|
Installation & monthly b |
| Spot check on Uniformity |
|
Daily b |
| Frequency Response | No resonant frequency components below 20 Hz a | Installation & yearly b |
a
As recommended by the Association de Française (Bizzo et
al. 1985).
b
Recommended from experimental evidence of the stability of a prototype
force platform evaluated over a year (Browne 1999).
Additional
information about the stability of the individual transducers
can be obtained if the output voltage from the ADC is accessible.
The stability over time of the transducers can be determined by
applying a calibrated load between 10 and 40 kg directly over
each of the transducers and measuring the output voltage response
of the transducers from the analogue to digital converter. It
is important that the load is accurately placed in the same position
each time: this can be achieved by using a placement mat as described
in the uniformity test procedure.
The Association
Française de Posturologie made no recommendations for the
temporal stability of COP measurements; however, upon consideration
of the small magnitude of difference, 2 mm, between patients with
Parkinson's disease and age-matched healthy subjects it is evident
that COP measurements should also be precise to within 1 mm, in
order to detect changes in subject's ability to balance (Browne
1999, Berg 1989, Hufschmidt et al. 1980).
This test
should be carried out at a minimum of monthly intervals unless
the stability of the system is such that it needs to be carried
out more frequently.
2.1.2. Dynamic test of the QC procedure.
The
frequency response of a force
platform can be determined by using a rocking device such as a
pendulum, which would simulate the sway of a subject, and measuring
the frequency and amplitude of the rocking device, using the force
platform's own processing and analysis program. The rocking device
frequency and amplitude would need to be controllable and be able
to simulate a range of frequencies from 0-20 Hz. Therefore, it
could be evaluated whether the force platform reproduces all the
input frequencies and at the correct amplitude.
Additional
information about the force platform's frequency response can
be obtained if the output voltage from the ADC is accessible.
The frequency response of the force platform can be determined
by applying an impulse to the force platform (Bizzo et al.
1985, Starck et al. 1993, Hall et al. 1996). An
impulse can be applied by dropping a calibrated load between 5
and 10 kg onto the top plate of the force platform. Once an impulse
has been applied to the force platform the point spread function
(PSF) of the system can be measured from the output voltage of
the analogue to digital converter.
|
The fast Fourier transform of the PSF can then be calculated and from this the modulation transfer function for the system can be determined. The natural frequency of the system can be determined from either the time domain plot of the point spread function or from the frequency domain plot of the point spread function. The modulation transfer function will indicate how the system represents the frequency components of sway. The range of frequency components in sway is between 0 and 10 Hz, therefore the recommended frequency of force platforms is that the natural frequency of the force platform should not be lower than 20 Hz (Bizzo et al. 1985). If the force platform has frequency components lower than 20 Hz then the amplitude of the subject's frequency components of their sway may be increased or decreased. |
FIG. 4 - Point Spread Function |
This test
should be carried at acceptance testing and at a minimum of yearly
intervals thereafter unless the stability of the system is such
that it needs to be carried out more frequently.
A QC
procedure has been described which enables users to determine
whether a force platform is performing to the required specifications,
as provided by the Association Française de Posturologie
(Bizzo et al. 1985). The present paper is the first one,
to the author's knowledge, which outlines a comprehensive QC procedure
for force platforms and which recommends appropriate time intervals
for carrying out each of the tests in the QC procedure. However,
the time intervals suggested are only guidelines established after
a year of testing a prototype force platform and, therefore, further
work needs to be carried out in order to establish a baseline
of time intervals for force platforms of varying ages and varying
complexity.
From the
results of the QC procedure sources of error caused by the electronic
components of the force platform can be identified and if the
output voltage from the ADC is accessible then the exact source
of the error can be identified for example excessive nonlinearity
of the transducer response, different offset voltages of the transducers
or noise due to the ADC or mains. Corrections can be applied to
correct for excessive nonlinearity and different offset voltages
in the transducers. Noise due to the ADC can be reduced by using
a higher sampling frequency and then using a low pass filter,
while noise from the mains can be reduced with better insulating
shielding on the force platform wiring.
In this
paper a QC test procedure is outlined to allow the performance
of force platforms to be evaluated and monitored over time. This
QC test procedure includes details of what QC tests should be
carried out; how the QC tests may be carried out; the frequency
with which they should be carried out and the typical performance
required in order for the force platform's measurements to be
regarded as accurate, precise and giving a faithful reproduction
of the frequency components of the subject's sway. Without these
QC tests being carried out it is difficult for investigators to
trust the results obtained and to compare results obtained for
a patient undergoing physiotherapy as, the results obtained may
not be accurate or precise. Table 1 is a summary of the QC protocol.
References